APL Valve
Adjustable Pressure Limitation valve (APL)
Selecting knob for »MAN« and »SPONT« on pressure limiting:
The APL valve has two functions. It limits the maximum pressure during manual ventilation. It also exhausts excess gas into the scavenger system during manual and spontaneous ventilation.
The APL valve is connected to the patient airway through the ventilator. It functions only when the ventilator is in ManSpont mode or ventilator override condition.
The APL valve has a labeled knob for selecting between spontaneous and manual modes of ventilation and for indicating approximate pressure settings.
NOTE:
The APL valve is automatically excluded from the breathing circuit whenever anautomatic ventilation mode is selected.